1 2 3
Description
3
1 Description
TRIVAC-B pumps are oil-sealed rotary vane pumps. The
TRIVAC S16 B and S 25 B are single-stage pumps, and
the TRIVAC D16 B and D 25 B are dual-stage pumps.
The number in the type designation (16 or 25) indicates
the pumping speed in m
3
· h
-1
.
TRIVAC-B pumps can pump gases and vapours and
evacuate vessels or vacuum systems in the fine vacuum
range. Those of standard design are not suitable for
pumping greater than atmospheric concentrations of
oxygen, hazardous gases, or extremely aggressive or
corrosive media.
The drive motor of the TRIVAC-B is directly flanged to
the pump at the coupling housing. The pump and motor
shafts are directly connected by a flexible coupling. The
bearing points of the pump module are force lubricated
sliding bearings. All controls as well as the oil-level glass
and the nameplate are arranged on the front. All connec-
tions are to be found at the sides of the pump. The oil-
level glass is provided with prisms for better observation
of the oil level.
The pump module consists of assembly parts which are
pin-fitted so as to allow easy disassembly and reassem-
bly. The pump module can be easily removed without
special tools.
1.1 Function
The rotor (2/7), mounted eccentrically in the pump hou-
sing (2/6), has two radially sliding vanes (2/5) which divi-
de the pump chamber into several compartments. The
volume of each compartment changes periodically with
the rotation of the rotor.
As a result, gas is sucked in at the intake port (2/1). The
gas passes through the dirt trap sieve (2/2), flows past
the open anti-suckback valve (2/3) and then enters the
pump chamber (2/6). In the pump chamber, the gas is
passed on and compressed, after the inlet aperture is
closed by the vane.
The oil injected into the pump chamber is used for sea-
ling and lubricating. The slap noise of the oil in the pump
which usually occurs when attaining the ultimate pressu-
re is prevented by admitting a very small amount of air
into the pump chamber.
The compressed gas in the pump chamber is ejected
through the exhaust valve (2/10). The oil entrained in the
gas is coarsely trapped in the internal demister (2/11);
there the oil is also freed of mechanical impurities. The
gas leaves the TRIVAC-B through the exhaust port.
During compression, a controlled amount of air - the so-
called gas ballast - can be allowed to enter the pump
chamber by opening the gas ballast valve. The gas bal-
last stops condensation of vapours in the pump chamber
up to the limit of water vapour tolerance as specified in
the technical data for the pump.
Key to Fig. 1
1 Oil filter OF 4-25
2 Exhaust filter AF 16-25
3 Condensate trap AK 16-25
Fig. 1 TRIVAC-B with accessories